HSA/FSA-eligible products available
Medical Supplies for Home
#1011073254 | Mfg #PM100N-HC | Diagnostic Instruments and Supplies

Pulse Co-Oximeter Nellcor™

OXIMETER, PULSE BEDSIDE
Pulse Co-Oximeter Nellcor™

Incorporates the latest Nellcor™ digital signal processing technology for accurate, reliable readings even during low perfusion, motion and other forms of signal interference

Product Specifications

MSFH #
1011073254
Manufacturer #
PM100N-HC
Brand
Nellcor™
Manufacturer
Medtronic MITG
Country of Origin
Unknown
Alarm Type
Audible/Visual
Application
Pulse Co-Oximeter
HCPCS
E0445
Power Source
Battery Operated
UNSPSC Code
42181805

Description

Provides clinicians with real‑time information regarding their patients’ respiratory status, including continuous SpO2 and pulse rate monitoring and trending data

Includes SatSeconds alarm management, a clinician‑controlled feature that can distinguish between real, clinically significant events and transient events by taking into account both the severity and the duration of any desaturation event

Meets IEC 60601-1-11 standards for home health equipment compliant homecare and sleep study modes for safe and effective use of the monitor by lay users and in non‑hospital settings

1% to 100% SpO2, 20 to 250 beats per minute (bpm) pulse rate; 0.03% to 20% pulse amplitude measurement range

Saturation - adult: 70% to 100% ± 2 digits; adult and neonate low sat: 60% to 80% ± 3 digits; neonate: 70% to 100% ± 2 digits; low perfusion: 70% to 100% ± 2 digits; adult and neonate with motion: 70% to 100% ± 4 digits

Pulse rate - adult and neonate: 20 to 250 bpm ± 3 digits; low perfusion: 20 to 250 bpm ± 3 digits; adult and neonate with motion: 20 to 250 bpm ± 5 digits

Power requirements: 100 to 240 V AC, 50/60 Hz, 45 VA; Fast-acting 2 A 32 V AC/DC, fast-acting 500 mA 32 V AC/50 DC fuse rating; lithium ion battery; minimum of five hours using new, fully charged battery with no alarms; optional 10-hour battery capacity

Operating temperature - Instrument: 5 to 40 °C; Transport/storage temperature: 20 to 60 °C

Operating humidity - 15% to 93% noncondensing

-170 to 4877 Meter Operating altitude

82 X 155 X 255 mm Size

Accuracy - Accurately assesses patients’ status with pulse oximetry measurements of ± 2 for 70% to 100% saturation, and low saturation accuracy of ± 3 for 60% to 80%

Speed - reacts to patient status with technology that displays patient oxygenation and pulse rate more quickly than other technologies

Motion tolerance - Accurately assesses patients’ status during periods of movement or noise, avoiding dropouts or delays, Covidien is the first company to receive FDA clearance for a motion tolerant pulse oximeter that is also compliant with ISO 80601-2-61

Displays real time SpO2 and pulse rate measurements, plethysmographic waveforms and pulse amplitude

SatSeconds alarm management

Sleep Study mode

Customer Reviews

No reviews yet. Be the first to review this product.

Write a Review