DIAGNOSTIC SYY, LAB GENEXPERT I W/LAPTOP
1 Reaction site
Product Specifications
Description
Fully integrated and automated single module on-demand molecular diagnostic system
Single module instrument
Heating ramp rates (max.): 10°C/sec from 50 to 95°C
Cooling ramp rates (max.): 2.5°C/sec from 95 to 50°C
Temperature duration accuracy: ± 1.0 sec From programmed time
Temperature accuracy: ± 1.0°C from 60 to 95°C
Melt curve programmable ramp rates: 0.01°C/seconds to 1.0°C/seconds
The GeneXpert® system fully integrates and automates sample extraction, amplification and detection — all in one cartridge